Brand Name : LABNOVATION
Model Number : LD-560
Certification : IFCC, NGSP, CE
Place of Origin : Shenzhen, China
MOQ : 1 Piece
Price : US$5000-12000/Piece
Payment Terms : D/A, D/P, T/T
Supply Ability : 100 Units/Day
Delivery Time : 5-7 work day
Packaging Details : Carton
Method : HPLC
Test Speed : 90 seconds/test
Sample Type : Whole Blood, Pre-diluted blood
Column Capacity : 3000 Test
Dimensions : 378*380*510mm
Certificated : NGSP,IFCC
Data Storage : 5000 Test
Precision : CV<2%
Labnovation LD-560 HbA1c Analyzer HPLC System For HbA1c Testing Dual Mode With NGSP&IFCC Certificated
Intended Use
The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin (HbA1c) in human blood.
Specification
Model | LD-560 |
Method | HPLC |
Parameters | HbA1c, HbF, HbA1a, HbA1b, HbA0; HbA2 (optional) |
Test Speed | 90 seconds per test |
Reagents kit | Eluent A, Eluent B, Hemolysis |
Autoloader Capacity | 20 tubes |
Precision | CV<2% |
Dimension | 378(L) ×380(W)×510(H)mm |
Weight | 28kg |
Full automatic smart system
Automatic scanning of barcode
Autoloader for 20 tubes
Easy to change reagents
Operation Preparation
Before operating this analyzer, please read the operation manual carefully to avoid malfunction or damage.
If the analyzer is used in a manner not specified by the manufacturer, the protection provided by the analyzer may be destroyed.
Before turning on the analyzer, the operator must check the following requirements to ensure that the system is ready.
(1) Check the remaining amount of reagents
Check whether the remaining amount of eluent A, eluent B, eluent C and hemolysis agent is sufficient.
(2) Check the waste liquid barrel
Check whether the waste liquid barrel has been emptied before starting the analyzer every day.
(3) Check the liquid tubes and power supply
Check whether the tubes of reagent and waste liquid are bent and whether the connections are reliable.
Check the analyzer for fluid leakage.
Check whether the power plug of the analyzer is securely inserted into the power socket.
(4) Check the printer
Check whether the paper roll is sufficient and whether it is installed in correct place.