Sign In | Join Free | My frbiz.com |
|
Brand Name : LABNOVATION
Model Number : LD-560
Certification : IFCC, NGSP, CE
Place of Origin : Shenzhen, China
MOQ : 1 Unit
Price : USD7,000.00-12,000.00/Unit
Supply Ability : 100 Units/Day
Delivery Time : 15-20 working days
Packaging Details : Carton Size: 480mm×480mm×615mm, Gross Weight: 28kgs
Test Method : HPLC (High Performance Liquid Chromatography)
Test Time : 90s/test
CV Precision : <2%
Sample types : Whole blood or Pre-diluent sample
Test blood volume : 5μL in whole blood mode; 150μL in pre-diluted mode
Sample capacity : 20 samples
Reagent pack : 600T
Column test quantity : 3000T
Automatic HPLC Glycosylated Hemoglobin HbA1c Analyzer for Diabetes Quantitative Detection
The Fully Automated HbA1c Analyzer LD-560 is used for the quantitative detection of glycosylated hemoglobin (HbA1c) in human blood.
It offers the fast throughput of HbAlc results in 90 seconds, with Hb variant detection, providing the outstanding solution for quick and reliable diabetic monitoring. No sample preparation and very little hands-on time by the operator is required for the LD-560 Analyzer.
Specification
Model | LD-560 |
Methodology | High-Performance Liquid Chromatography (HPLC) |
Test Modes | Fast Mode, Variant Mode |
Test Range | 3% - 18% |
Precision (CV) | <2% |
Parameter | HbA1c, HbF, HbA1a, HbA1b, HbA0, HbA2(Optional) |
First Sample Result | 4 Mins |
Test Speed | Fast Mode: 90 Seconds/Test Variant Mode: 140 Seconds/Test |
Sample Type | Whole blood, Pre-diluted blood |
Sample Volume | Whole blood: 5μL Pre-diluted blood: 150μL |
Sample loading capacity | 20 |
Photometer | 420nm |
Column Capacity | 3000T |
Filter Capacity | 600T |
Display | 10“ LCD touch screen |
Reagent Kit | Eluent A, Eluent B, Eluent C, Hemolysis, Chromatography Column |
Storage | 5000 Sample Results |
Connection interface | USB, LAN, LIS Compatible |
Printer | Thermal Printer |
System operation environment | Temperature 15-30℃ Humidity 30%-85% |
Power | AC 100-240V 50/60HZ 300VA |
Dimensions | 378mmx380mmx510mm |
Weight | 28Kgs |
Features
*HPLC, CV<2% - Accurate and Stable
*Full Automatic smart System
*Quality & Performance - IFCC & NGSP Certificate
*User Friendly
Main Parts
The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.
Reagents and Consumables
Eluent A/Eluent B/Eluent C(Variant Mode) | Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin. |
Hemolysis Reagent | Used to dissolve red blood cells and release hemoglobin. |
Chromatography Column | Used for separation and elution of glycosylated hemoglobin and non glycated. |
Hemoglobin Filter | Used to initially filter the sample after hemolysis. |
Paper Rolls | Prints a report of the test results. |
Certificate
![]() |
Automatic HPLC Glycosylated Hemoglobin A1c Analyzer For Diabetes Quantitative Detection Images |